The Common problem (faults) and solution with intense pulsed light(IPL) treatment
Intense pulsed light therapy machine (using the 800W and 2000W power supply)
1, cant light the lamp of ipl handle;
Failure analysis:
(1)ipl handpiece(ipl lamp) bad
(2) bad power supply
(3)simmer board failure
Solution: Try to replace a new ipl handle (IPL lamp), in case the problem
is still same ,replace the ipl power
2, no effect after treatment, the treatment energy is too low ;
Failure analysis: (1) ipl handle used for a long time, energy consumption of ipl xenon lamp.
(2) power supply or LCD is not working properly, the voltage of the capacitor is low.
Solution: (1) explain the situation to the customer, make the appropriate adjustments to the operating parameters, serious energy consumption of IPL xenon lamp, the longevity of ipl lamp is use up .
Need to replace a new ipl lamp or ipl handle
(2) replace the power supply or LCD.
3, unstable energy or energy become weak gradually , and finally no light shot;
Failure analysis: (1) .
Each time a larger difference between the discharge interval, the pulse interval is shorter.
(2) Power , LCD control panel not work
Solution: (1) to keep each discharge interval of 3 seconds.
(2) replace the power supply or LCD.
4, cant turn ipl machine on ;
Failure analysis: (1) bad power cord
(2) Emergency switch does not bounce
(3) key switch bad
(4) Bad fuse (burn up fuse)
(5) the air switch is disconnected
Solution: (1) the replacement a new power cord
(2) Emergency switch bounce
(3) Replace key switch
(4) Replace the fuse
(5) the air switch to closed state
5, ipl shot automatically after simmer ;
Failure Analysis: IGBT bad
Solution: Replace the broken IGBT or replace the power , need professional maintenance(Beijing ).
6, counting ,but no light (no ipl shot );
Failure Analysis: LCD panel or power board bad
Solution: replace the power supply or LCD panel
7, ipl handle heat
Failure analysis: (1) there is no water flow ,or no water added in the water tank
(2) long working hours
Solution: (1) adding water to the overflow hole, place the ipl handle on the ground, so as to make sure that the water circulation is good enough
(2) Do not work too long. For a lot of clients , pay attention to changing the water, first water tank release 2 / 3, keep 1 / 3, re-add water to overflow hole, rest 30-40 minutes before work.
8, touch screen displays only half or blue screen;
Failure Analysis: Bad LCD
Solution: Replace LCD
9, LCD screen blank, pump and fan working properly;
Failure analysis: (1) LCD bad
(2) no power supply for LCD screen
Solution: (1) the replacement of the LCD
(2) check the cable, to ensure that the LCD's main power supply.
10, automatic lighting(ipl lamp light automatically );
Failure analysis: (1) LCD bad
Solution: (2) the replacement of the LCD
11, showing no water circulation or no water;
Failure analysis: (1) bad water pump
(2) the short circuit for water probe
(3) LCD panel bad or power supply bad
Solution: (1) take out the ipl handle, if there is flowing out, indicating that the pump normally, if there is no water in the ipl handle . water pump bad,
You can also hear the sound of water flow on the ipl handle when you turn on machine
(2) connect water probe well.
(3) the replacement of the LCD panel or the power board
12, ipl handle is not cooling, not shot, room temperature;
Failure analysis: (1) switching power supply bad (12V DV switch power)
(2) Refrigeration signal cable is not plugged in
(3) cooling chip bad
Solution: (1) the replacement of switching power supply
(2) connect the refrigeration signal cable well.